Abstract

The aim of this work is the analysis of the contact area in a knee prosthesis using two different approaches. In particular, the interface between the femoral component and the polyethylene insert has been studied both numerically and experimentally. The interest in studying the contact area is related to the fact that the wear of the polyethylene insert, due to the high contact pressures, represents one of the major causes of failure of the total knee prosthesis. The possibility to evaluate the contact area at different loads and mutual position between femur and tibia is, therefore, of fundamental importance to study the service life of a prosthesis and to improve its performance. The finite element numerical approach has required the acquisition, through reverse engineering, and CAD modelling of the prosthetic components. Then the FEM simulations have been developed considering two different load conditions. In order to compare the calculated data, the same load configurations have been used for experimental tests based on ultrasonic method. In this case, some preliminary tests were required to calibrate the system depending on the particular characteristics of materials, geometries and surface finish of the prosthesis.The results show a good correlation between the data obtained with the two different approaches and, consequently, a good level of reliability of the procedures developed for the numerical and experimental evaluation of the contact area. The numerical procedure can be used to determine the area for different angles and loads, but especially in the design phase. The ultrasonic technique can be used to validate the numerical data.

Article history

  1. ReceivedUntil the next stage: 0 days
  2. AcceptedUntil the next stage: 0 days
  3. Published

How to cite

APA

Cerniglia, D., Ingrassia, T., D’Acquisto, L., Saporito, M., & Tumino, D. (2012). Contact between the components of a knee prosthesis: numerical and experimental study. Fracture and Structural Integrity, 6(22), pages 56–68. https://doi.org/10.3221/IGF-ESIS.22.07

IEEE

D. Cerniglia, T. Ingrassia, L. D’Acquisto, M. Saporito, and D. Tumino, “Contact between the components of a knee prosthesis: numerical and experimental study,” Fracture and Structural Integrity, vol. 6, no. 22, pp. pages 56–68, 2012, doi: 10.3221/IGF-ESIS.22.07.

MLA

Cerniglia, D., et al. “Contact between the components of a knee prosthesis: numerical and experimental study.” Fracture and Structural Integrity, vol. 6, no. 22, 2012, pp. pages 56–68. https://doi.org/10.3221/IGF-ESIS.22.07.

Chicago

Cerniglia, D., T. Ingrassia, L. D’Acquisto, M. Saporito, and D. Tumino. 2012. “Contact between the components of a knee prosthesis: numerical and experimental study.” Fracture and Structural Integrity 6 (22): pages 56–68. https://doi.org/10.3221/IGF-ESIS.22.07.

Harvard

Cerniglia, D., Ingrassia, T., D’Acquisto, L., Saporito, M. and Tumino, D. (2012) ‘Contact between the components of a knee prosthesis: numerical and experimental study’, Fracture and Structural Integrity, 6(22), pp. pages 56–68. doi: 10.3221/IGF-ESIS.22.07.

Vancouver

Cerniglia D, Ingrassia T, D’Acquisto L, Saporito M, Tumino D. Contact between the components of a knee prosthesis: numerical and experimental study. Fracture and Structural Integrity. 2012;6(22):pages 56-68. doi:10.3221/IGF-ESIS.22.07

BibTeX

@article{cerniglia2012, author = {Cerniglia, D. and Ingrassia, T. and D’Acquisto, L. and Saporito, M. and Tumino, D.}, title = {Contact between the components of a knee prosthesis: numerical and experimental study}, journal = {Fracture and Structural Integrity}, year = {2012}, volume = {6}, number = {22}, pages = {pages 56--68}, doi = {10.3221/IGF-ESIS.22.07}, url = {https://doi.org/10.3221/IGF-ESIS.22.07}, issn = {1971-8993} }

RIS

TY - JOUR AU - Cerniglia, D. AU - Ingrassia, T. AU - D’Acquisto, L. AU - Saporito, M. AU - Tumino, D. TI - Contact between the components of a knee prosthesis: numerical and experimental study JO - Fracture and Structural Integrity T2 - Fracture and Structural Integrity PY - 2012 DA - 2012/09/30 VL - 6 IS - 22 SP - pages 56 EP - 68 DO - 10.3221/IGF-ESIS.22.07 UR - https://doi.org/10.3221/IGF-ESIS.22.07 SN - 1971-8993 ER -

Licence

Copyright © 2014 Frattura ed Integrità Strutturale.

Copyright 
Authors are allowed to retain both the copyright and the publishing rights of their articles without restrictions.  

Open Access Statement

Fracture and Structural Integrity (F&SI) is an open-access journal which means that all content is freely available without charge to the user or his/her institution. Users are allowed to read, download, copy, distribute, print, search, or link to the full texts of the articles in this journal without asking prior permission from the publisher or the author. This is in accordance with the DOAI definition of open access.

F&SI operates under the Creative Commons Licence Attribution 4.0 International (CC-BY 4.0). This allows to copy and redistribute the material in any medium or format, to remix, transform and build upon the material for any purpose, even commercially, but giving appropriate credit, providing a link to the license, and indicating if changes were made.

Similar articles